Thermodynamic interpretation of the drift and noise of gradient baselines in reversed-phase liquid chromatography using mobile phase additives

Journal

JOURNAL OF CHROMATOGRAPHY A
Volume 1633, Issue -, Pages -

Publisher

ELSEVIER
DOI: 10.1016/j.chroma.2020.461605

Keywords

RPLC gradient baseline drift and noise; Acetonitrile-water gradients; Mobile phase additive; Trifluoroacetic acid; Equilibrium-dispersive model of; chromatography; Numerical calculations

Ask authors/readers for more resources

The drift and noise of acetonitrile-water gradient baselines (5-95%, v/v, 5 min linear gradient) in reversed phase liquid chromatography (RPLC) are recorded at a wavelength of 215 nm using 0.1% trifluoroacetic acid (TFA) as the mobile phase additive, a 4.6 mm x 150 mm 5 mu m Symmetry-C 18 RPLC column, and an Arc system (low-pressure gradient proportioning valve or GPV, pump with a stroke volume of either 66 or 132 mu L, no mixer) as the LC instrument. These observations are predicted from solid-liquid adsorp-tion thermodynamics which requires the measurement of the excess adsorption isotherm of acetonitrile from water onto the RPLC column and of the variation of the Henry's constant of TFA as a function of the volume fraction of acetonitrile in the bulk mobile phase. The incomplete mixing of the acetonitrile and water packets delivered by the low-pressure GPV is represented by a sinusoidal perturbation of the programmed volume fraction of acetonitrile during the entire gradient. The variation of the TFA absorbance at 215 nm with increasing acetonitrile concentration is measured in order to transform TFA concentration into the observable absorbance unit. The drift and noise of the gradient baseline are calculated by solving numerically (Rouchon method) the equilibrium-dispersive (ED) mass balance equations of acetonitrile and TFA The agreement between the calculated and observed gradient baselines is very good as the proposed model of chromatography accurately accounts for the displacement of TFA between stationary and mobile phases (early excess and late deficit of TFA concentration relative to 0.1%) and for the frequency (equal to the ratio of the applied flow rate to the stroke volume) and the amplitude of the periodic noise recorded during the gradient. From a practical viewpoint, the drift of the gradient baseline can be minimized by maximizing the ratio of the gradient volume to the hold-up volume (10) and/or by minimizing the retention factor of the mobile phase additive in the water-rich eluent (k < 0.2). The reduction of the noise amplitude below 0.1 mAU as requested by the pharmaceutical industry imposes the ratio of the flow rate to the stroke volume of the pump to be larger than 1 Hz. This opens avenues towards the development of new GPV, pump, and mixers in order to mix efficiently the solvent packets delivered by conventional LC instrument. (c) 2020 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available