Inaccurate Glucose Sensor Values After Hydroxyurea Administration

Journal

DIABETES TECHNOLOGY & THERAPEUTICS
Volume 23, Issue 6, Pages 443-451

Publisher

MARY ANN LIEBERT, INC
DOI: 10.1089/dia.2020.0490

Keywords

TPIAT; Continuous glucose monitoring; Pediatrics; Diabetes; Pharmacology

Ask authors/readers for more resources

The study evaluated the impact of hydroxyurea (HU) administration on continuous glucose monitoring (CGM) systems in patients, finding a significant rise in sensor readings from 0.5-6 hours after administration. It is recommended that patients rely on glucometer testing during this period.
Objective: To assess the degree, duration, mean absolute relative difference (MARD), and error analysis of discrepant values per continuous glucose monitoring (CGM) systems after hydroxyurea (HU) administration. Research Design and Methods: Inpatient glucometer and CGM data from 16 total pancreatectomy/islet autotransplantation patients using Dexcom Professional G4 and 12 patients using Dexcom G6 were analyzed after daily dosing with HU. Timing of HU dosing and median of 9.5 days of sensor and glucometer values were assessed per patient. Results: A large positive elevation of sensor readings was identified after HU dosing. The greatest discrepancy between glucometer and sensor readings occurred 0.5-2 h after HU administration [G4 (mean 3.0 mmol/L, median 2.4 mmol/L, MARD 55%), G6 (mean 4.2 mmol/L, median 4.6 mmol/L, MARD 91%)]. The discrepancy was <1.1 mmol/L, mean (-0.5 mmol/L) and median (-0.5 mmol/L), MARD 14% (G4) and <1.1 mmol/L, mean (0.3 mmol/L) and median (0.3 mmol/L), MARD 17% (G6), by 6 h after administration. Error analysis with the G6 system found 94% of pairs in clinically acceptable range by 6-9 h after HU administration. Aspirin, also given once daily, did not result in glucose value discrepancy with the G6 system but variability was observed with the G4 system. Conclusions: There was marked elevation of sensor glucose readings compared with glucometer values [up to 13.9 mmol/L (G4), 13 mmol/L (G6)] from 0.5 to 6 h after HU administration. It is important to counsel a patient using a Dexcom CGM system and HU therapy on this finding and to advise reliance on glucometer testing for accurate glucose assessment up to 6-9 h after HU administration.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available