4.1 Article

Hand-held dynamometer to measure pressure pain thresholds: A double-blinded reliability and validity study

Journal

MUSCULOSKELETAL SCIENCE AND PRACTICE
Volume 51, Issue -, Pages -

Publisher

ELSEVIER
DOI: 10.1016/j.msksp.2020.102268

Keywords

Algometry; Pain mechanisms; Quantitative sensory testing; Spatial summation

Categories

Ask authors/readers for more resources

This study aimed to validate the use of a hand-held dynamometer (HHD) as a tool for measuring pressure pain threshold (PPT). The results showed that when force was normalized by the circumference of the applicator, the HHD demonstrated good reliability and validity in measuring PPT. Clinicians may consider using the HHD to detect relevant pain mechanisms in their evaluation and treatment of pain.
Background: Pain is the most common complaint reported in the musculoskeletal setting. Quantitative sensory testing (QST) assists with pain mechanism identification, although QST is typically performed in research settings. It is possible that clinical utilization of QST may improve if clinically accessible tools can be reliably and validly used. Objective: To determine if a hand-held dynamometer (HHD) can be a valid and reliable assessment of pressure pain threshold (PPT). Design: Double-blinded validation study. Methods: Eighteen healthy subjects (25.6 +/- 3.4 years old) participated in this study. Two testers independently assessed PPT using a HHD and a digital algometer. Assessments followed previously described pressure algometry protocols. Testers and subjects were each blinded to data during assessments. Results: Intra- and inter-rater reliability were excellent for the foot and face for both devices (ICC' s > 0.9). Bland-Altman plots and intraclass correlation coefficients revealed good-excellent agreement with minimal proportional bias when normalizing device force at pain threshold to the circumference of the device applicator (ICC 95%CI: 0.56-0.95). Only poor-good agreement (ICC 95% CI: 0.30-0.76) and significant proportional bias was observed when normalizing to area (pressure). Conclusions: Based on the results of this study, when force is normalized by circumference of the applicator, a HHD was found to be a valid and reliable tool for measuring PPT. Clinicians may use HHD to detect relevant pain mechanisms at fault in their evaluation and treatment of pain. Additional research in various pathologic populations is warranted.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.1
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available