4.6 Article

Comparison of Low Dose Performance of Photon-Counting and Energy Integrating CT

Journal

ACADEMIC RADIOLOGY
Volume 28, Issue 12, Pages 1754-1760

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.acra.2020.07.033

Keywords

Photon counting; Computed tomography; Low dose; Image quality

Funding

  1. National Institutes of Health (NIH) Clinical Center Radiology and Imaging Sciences (RADIS)
  2. National Institute of Biomedical Imaging and Bioengineering (NIBIB), NIH Graduate Partnership Program
  3. NIH Intramural Research Program [NIH Z01 1ZID BC011242, CL040015]
  4. Siemens Medical Solutions

Ask authors/readers for more resources

The study showed that PCCT has superior image quality compared to EID CT at low doses, with significant improvements in image noise, contrast-to-noise ratio, and detectability index. Additionally, PCCT demonstrated higher CNR for iodine contrast at all dose levels and iodine concentration levels.
Rationale and Objectives: The purpose of this study was to investigate the potential of photon-counting CT (PCCT) to improve quantitative image quality for low dose imaging compared to energy-integrating detector CT (EID CT). Materials and Methods: An investigational scanner (Siemens, Germany) with PCCT and EID CT subsystems was used to compare image quality performance at four dose levels: 1.7, 2, 4, 6 mGy CTDIvol, all at or below current dose values used for conventional abdominal CT. A CT quality control phantom with a homogeneous section for noise measurements and a section with cylindrical inserts of air (-910 HU), polystyrene (50 HU), acrylic (205 HU), and Teflon (1000 HU) was imaged and characterized in terms of noise, resolution, contrast-to-noise ratio (CNR), and detectability index. A second phantom with a 30 cm diameter was also imaged containing iodine solutions ranging from 0.125 to 8 mg I/mL. CNR of the iodine vials was computed as a function of CT dose and iodine concentration. Results: With resolution unaffected by dose in both PCCT and EID CT, PCCT images exhibited 22.1-24.0% improvement in noise across dose levels evaluated. This noise improvement translated into a 29-41% improvement in CNR and 20-36% improvement in detectability index. For iodine contrast, PCCT images had a higher CNR for all combinations of iodine contrast and dose evaluated. Conclusion: For the conditions studied, PCCT exhibited superior image quality compared to EID CT. For iodine detection, PCCT offered a notable advantage with improved CNR at all doses and iodine concentration levels.