Performance of Different Active Humidification Systems in High-Flow Oxygen Therapy

Journal

RESPIRATORY CARE
Volume 65, Issue 9, Pages 1250-1257

Publisher

DAEDALUS ENTERPRISES INC
DOI: 10.4187/respcare.07654

Keywords

high-flow oxygen therapy; active humidification; oxygen; intensive care unit

Ask authors/readers for more resources

BACKGROUND: We sought to evaluate the performance in terms of absolute humidity (AH), relative humidity (RH), and temperature of different heated humidifiers (HH) and circuits that are commonly used to deliver high-flow oxygen therapy in conventional ranges (30-60 L/min) and unconventional ranges (70-100 L/min). METHODS: In this prospective, observational study, an electronic thermohygrometer was used to obtain the required measurements. A mechanical ventilator was used as a source for high-flow nasal cannula oxygen therapy. For active humidification, the following equipment was used: a HH with standard disposable water trap circuit, 3 servo-controlled HH, and 7 circuits with a heated wire. Data on environmental conditions (ie, temperature, RH, AH) were collected from the laboratory during each measurement; the temperature, RH, and AH resulting from the application of 8 flows (30-100 L/min) were also recorded. Variables were compared with analysis of variance for repeated measurements with Tukey post hoc tests. A value of P < .05 was assumed to be significant. RESULTS: During the study, a statistically significant difference was found in the average AH for each flow for the different devices (P < .005). The highest AH values were recorded with the Fisher & Paykel MR850 and the Medtronic-DAR circuit (AH = 40.8 mg/L with flow of 50 L/min, P < .005), and the lowest AH values were recorded with the Flexicare FL9000 HH and the Hexicare circuit (AH = 11.4 mg/L with 100 L/min flow, P < .005). For flows > 50 L/min, the best performance for all flows in terms of AH was found with the Fisher & Paykel MR850 HH, regardless of the circuit used. CONCLUSIONS: During oxygen therapy with very high gas flows, HH devices behave differently and in many cases are inefficient in delivering adequate humidification, even at conventional flows. Caution is therefore recommended when selecting the device and flow settings for the implementation of high-flow nasal cannula oxygen therapy.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available