4.0 Article

Continuous blood pressure measurement using the pulse transit time: Comparison to intra-arterial measurement

Journal

BLOOD PRESSURE
Volume 24, Issue 4, Pages 217-221

Publisher

TAYLOR & FRANCIS LTD
DOI: 10.3109/08037051.2015.1030901

Keywords

Blood pressure; pulse transit time; validation

Ask authors/readers for more resources

Continuous blood pressure (BP) measurement allows the investigation of transient changes in BP and thus may give insights into mechanisms of BP control. We validated a continuous, non-invasive BP measurement based on the pulse transit time (PTT), i.e. BP (PTT), by comparing it with the intra-arterial BP (BP (i.a.)) measurement. Twelve subjects (five females and seven males) were included. BPi.a. was obtained from the radial artery using a system from ReCor Medical. Systolic and diastolic BP were calculated using the PTT (BPPTT, SOMNOscreen (TM)). PTT was determined from the electrocardiogram and the peripheral pulse wave. The BP was modulated by application of increasing doses of dobutamine (5, 10, 20 mu g/kg body mass). Systolic BP PTT and systolic BP i.a. correlated significantly (R = 0.94). The limits of agreement in the Bland-Altman plot were +/- 19 mmHg; the mean values differed by 1 mmHg. The correlation coefficient for the diastolic BP measurements was R - 0.42. The limits of agreement in the Bland-Altman plot were +/- 18 mmHg, with a mean difference of 5 mmHg in favour of the BP PTT. The study demonstrates a significant correlation between the measurement methods for systolic BP. The results encourage the application of PTT-based BP measurement for the evaluation of BP dynamics and pathological BP changes.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.0
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available