Evaluation of In Vitro Wettability of Soft Contact Lenses Using Tear Supplements

Journal

EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE
Volume 47, Issue 5, Pages 244-248

Publisher

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/ICL.0000000000000698

Keywords

Wettability; Contact lens; Contact angle; Hydrogel lens; Silicone hydrogel lens

Categories

Funding

  1. JSPS KAKENHI [JP15K10882, JP19K09961]

Ask authors/readers for more resources

The measured contact angles (CAs) on soft contact lenses (SCLs) were affected by the use of 2-MPC solutions and hyaluronate acid (HA), with silicone hydrogel SCLs showing higher CAs. The interaction between tear supplements and SCLs may vary based on ingredient agents, surface treatment of the lenses, and components of the supplements.
Objective: To evaluate the interaction between tear supplements and soft contact lenses (SCLs), we measured the contact angles (CAs) on the SCLs using commercially available tear supplements. Methods: We used four daily disposable conventional hydrogel lenses (etafilcon A, etafilcon A+ polyvinylpyrrolidone, nelfilcon A, and omafilcon A containing 2-methacryloyloxyethyl phosphorylcholine [MPC]) and four silicone hydrogel lenses (narafilcon A, senofilcon A, delefilcon A, and stenfilcon A). The CAs on the SCLs were measured using a sessile drop technique and four different types of sessile drops, including saline, artificial tears, lubricants containing 2-MPC (MPC solution), and 0.1% hyaluronate acid (HA). Results: The CA values associated with the silicone hydrogel lenses were significantly (P<0.001) lower than those associated with the conventional hydrogel lenses with all four solutions. The mean CA of 0.1% HA was significantly (P<0.01) higher than that of saline. The mean CA of the MPC solution was significantly (P<0.01) lower than that of saline with the conventional hydrogel lenses but significantly (P<0.05) higher than that of saline with the silicone hydrogel lenses. Conclusions: The CAs associated with the silicone hydrogel SCLs were higher with the use of the MPC solutions and HA in vitro. The measured CAs may depend on ingredient agents, surface treatment of the CLs, and components of the tear supplements.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available