4.7 Article

Long-Term Safety and Efficacy in Continued Access Left Atrial Appendage Closure Registries

Journal

JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY
Volume 74, Issue 23, Pages 2878-2889

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jacc.2019.09.064

Keywords

atrial appendage; atrial fibrillation; LAAC; NVAF; stroke; Watchman

Ask authors/readers for more resources

BACKGROUND Long-term data on the safety and efficacy of left atrial appendage closure (LAAC) for stroke prevention in patients with nonvalvular atrial fibrillation remain limited. OBJECTIVES The purpose of this study was to evaluate 4.5- to 5-year data in 2 U.S. Food and Drug Association LAAC mandated registries (CAP[Continued Access to PROTECT-AF] and CAP2 [Continued Access to PREVAIL]) for safety and efficacy. METHODS Two registries of patients implanted with LAAC devices provide the largest source of follow-up data. Both accompanied their respective randomized clinical trials, PROTECT-AF (Watchman Left Atrial Appendage System for Embolic PROTECTion in Patients With Atrial Fibrillation) and PREVAIL (Prospective Randomized Evaluation of the WATCHMAN LAA Closure Device In Patients with Atrial Fibrillation versus Long Term Warfarin Therapy), which used the same endpoints (primary efficacy of composite of stroke, systemic embolism, cardiovascular/unexplained death, and safety). RESULTS CAP included 566 patients with an average follow-up of 50.1 months (2,293 patient-years), and CAP2 included 578 patients with an average follow-up of 50.3 months (2,227 patient-years). CAP2 patients were significantly older and had higher CHA(2)DS(2)-VASc (congestive heart failure, hypertension, age >= 75 years, diabetes mellitus, stroke/transient ischemic attack, vascular disease, age 65 to 74 years, sex category) scores (4.51 vs. 3.88; p < 0.001). Procedural success was similar in both (94%). The primary composite endpoint occurred at a rate of 3.05 per 100 patient-years in CAP and 4.80 per 100 patient-years in CAP2; events contributing to this endpoint were most commonly cardiovascular/unexplained death (1.69 per 100 patient-years for CAP and 2.92 per 100 patient-years for CAP2). Hemorrhagic stroke was significantly less than ischemic stroke (0.17 per 100 patient-years in CAP and 0.09 per 100 patient-years in CAP2), and total stroke rates were significantly less than predicted by CHA(2)DS(2)-VASc score (78% reduction with CAP, 69% reduction with CAP2). CONCLUSIONS These registries, which contain the longest and largest follow-up data of patients with the Watchman device, support LAAC as a safe and effective therapy for long-term anticoagulation in patients with nonvalvular atrial fibrillation, and document the lowest rate of hemorrhagic stroke identified in this population. (C) 2019 Published by Elsevier on behalf of the American College of Cardiology Foundation.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available