4.1 Article

Intervendor variability of carotid intima-media thickness measurement: validation study using newly developed ultrasound phantom

Journal

JOURNAL OF MEDICAL ULTRASONICS
Volume 47, Issue 2, Pages 155-165

Publisher

SPRINGER JAPAN KK
DOI: 10.1007/s10396-019-00995-7

Keywords

Intima-media thickness; Intervendor variability; Quality control

Ask authors/readers for more resources

Purpose Ultrasonography-derived carotid artery intima-media thickness (IMT) has been established as an early atherosclerotic imaging biomarker. The IMT reference value of a healthy person is approximately 0.1 x (every 10 years of age) + 0.2 (mm); accordingly, it requires an accuracy of at least 0.1 mm. However, one concern of IMT measurement is intervendor variability. In this study, we aimed to verify the intervendor variability using an IMT phantom. Methods An improved IMT phantom was developed, and it was possible to analyze the IMT by software for all vendors. Results With the vendor-specific software, the maximum difference between the devices was 0.08 mm, and the difference in quartile range was 0.06 mm. On the other hand, with the vendor-independent offline software, the maximum difference between the devices was 0.16 mm, and the quartile range of variation was 0.06 mm. Conclusion The intervendor variability assessed using our IMT phantom was less than 0.10 mm, and the on-board vendor-specific software was shown to reduce the difference between the devices significantly compared with the vendor-independent offline software. To further improve the vender difference, adjustment by means of vendor-specific software based on a standardized IMT phantom is warranted.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.1
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available