Optical imaging provides rapid verification of static small beams, radiosurgery, and VMAT plans with millimeter resolution

Journal

MEDICAL PHYSICS
Volume 46, Issue 11, Pages 5227-5237

Publisher

WILEY
DOI: 10.1002/mp.13797

Keywords

optical imaging; SBRT; scintillation; small fields; SRS

Funding

  1. NIBIB NIH HHS [R01 EB023909] Funding Source: Medline
  2. NIH HHS [R44 CA199681, R01 EB023909, P30 CA 23108, R01 EB024498] Funding Source: Medline

Ask authors/readers for more resources

PurposeWe demonstrate the feasibility of optical imaging as a quality assurance tool for static small beamlets, and pretreatment verification tool for radiosurgery and volumetric-modulated arc therapy (VMAT) plans. MethodsSmall static beams and clinical VMAT plans were simulated in a treatment planning system (TPS) and delivered to a cylindrical tank filled with water-based liquid scintillator. Emission was imaged using a blue-sensitive, intensified CMOS camera time-gated to the linac pulses. For static beams, percentage depth and cross beam profiles of projected intensity distribution were compared to TPS data. Two-dimensional (2D) gamma analysis was performed on all clinical plans, and the technique was tested for sensitivity against common errors (multileaf collimator position, gantry angle) by inducing deliberate errors in the VMAT plans control points. The technique's detection limits for spatial resolution and the smallest number of control points that could be imaged reliably were also tested. The sensitivity to common delivery errors was also compared against a commercial 2.5D diode array dosimeter. ResultsA spatial resolution of 1 mm was achieved with our imaging setup. The optical projected percentage depth intensity profiles agreed to within 2% relative to the TPS data for small static square beams (5, 10, and 50 mm(2)). For projected cross beam profiles, a gamma pass rate >99% was achieved for a 3%/1 mm criteria. All clinical plans passed the 3%/3 mm criteria with >95% passing rate. A static 5 mm beam with 20 Monitor Units could be measured with an average percent difference of 5.5 3% relative to the TPS. The technique was sensitive to multileaf collimator errors down to 1 mm and gantry angle errors of 1 degrees. ConclusionsOptical imaging provides ample spatial resolution for imaging small beams. The ability to faithfully image down to 20 MU of 5 mm, 6 MV beamlets prove the ability to perform quality assurance for each control point within dynamic plans. The technique is sensitive to small offset errors in gantry angles and multileaf collimator (MLC) leaf positions, and at certain scenario, it exhibits higher sensitivity than a commercial 2.5D diode array.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available