4.4 Article

High-Frequency Irreversible Electroporation Using 5,000-V Waveforms to Create Reproducible 2-and 4-cm Ablation Zones - A Laboratory Investigation Using Mechanically Perfused Liver

Journal

JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY
Volume 31, Issue 1, Pages 162-168

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jvir.2019.05.009

Keywords

-

Funding

  1. UNC/NCSU Joint Depatttnent of Biomedical Engineering

Ask authors/readers for more resources

Purpose: To investigate if high-frequency irreversible electroporation (H-FIRE) treatments can be delivered at higher voltages and with greater energy delivery rates than currently implemented in clinical irreversible electroporation protocols. Materials and Methods: Treatments using 3,000 V and 5,000 V were administered to mechanically perfused ex vivo porcine liver via a single applicator and grounding pad (A+GP) as well as a 4-applicator array (4AA). Integrated energized times (IET) 0.01-0.08 seconds and energy delivery rates 25-300 mu s/s were investigated. Organs were preserved at 4 degrees C for 10-15 hours before sectioning and gross analysis using a metabolic stain to identify the size and shape of ablation zones. Results: A+GP ablations measured between 1.6 cm and 2.2 cm, which did not increase when IET was increased from 0.02 seconds to 0.08 seconds (P > .055; range, 1.9-2.1 cm). Changes in tissue color and texture consistent with thermal damage were observed for treatments with energy delivery rates 50-300 mu s/s, but not for treatments delivered at 25 mu s/s. Use of the 4AA with a 3-cm applicator spacing resulted in ablations measuring 4.1-1.9 cm with energy delivery times of 7-80 minutes. Conclusions: H-FIRE treatments can rapidly and reproducibly create 2-cm ablations using an A+GP configuration. Treatments without thermal injury were produced at the expense of extended treatment times. More rapid treatments resulted in ablations with varying degrees of thermal injury within the H-FIRE ablation zone. Production of 4-cm ablations is possible using a 4AA.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available