High-quality, high-throughput cryo-electron microscopy data collection via beam tilt and astigmatism-free beam-image shift

Journal

JOURNAL OF STRUCTURAL BIOLOGY
Volume 208, Issue 3, Pages -

Publisher

ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.jsb.2019.09.013

Keywords

Cryo-EM data collection; Beam-image shift; Automatic coma determination program; Astigmatism

Funding

  1. National Key R&D Program of China [2017YFA0504700]
  2. Strategic Priority Research Program of CAS [XDB08030204]
  3. National Thousand (Young) Talents Program from the Office of Global Experts Recruitment in China

Ask authors/readers for more resources

The throughput of cryo-electron microscopy (cryo-EM) can be improved by employing a procedure that collects beam-image shift data. However, this procedure inadvertently induces a beam tilt, thus decreasing the resolution of the reconstruction. Here, we report an automatic calibration procedure for correcting the beam tilt and a large unexpected astigmatism in the beam-image shift data collection. In this procedure, the changes of the beam tilt and the astigmatism against the beam shift are measured and calibrated. The beam tilt and the astigmatism are corrected by changing the setting of the microscope using predicted values from the calibration. Using our corrected beam-image shift data collection, we found that the resolution remained identical as long as the distance of the beam shift was below 10 pm. The image throughput increases by 80%, with image quality improving by reducing the residual stage drift, thus benefiting the high resolution cryo-EM structure determination. Such a calibration procedure takes about 3 h and can be applied to different microscopes.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available