4.6 Article

A comparison of collision cross section values obtained via travelling wave ion mobility-mass spectrometry and ultra high performance liquid chromatography-ion mobility-mass spectrometry: Application to the characterisation of metabolites in rat urine

Journal

JOURNAL OF CHROMATOGRAPHY A
Volume 1602, Issue -, Pages 386-396

Publisher

ELSEVIER
DOI: 10.1016/j.chroma.2019.06.056

Keywords

Ion mobility spectrometry; Collision cross section; Metabolic phenotyping; Metabonomics; Metabolomics

Funding

  1. STRATIGRAD program at IC
  2. Marie Sklodowska-Curie EU Framework for Research and Innovation Horizon 2020 [675132]
  3. MRC [MR/N006321/1] Funding Source: UKRI

Ask authors/readers for more resources

A comprehensive Collision Cross Section (CCS) library was obtained via Travelling Wave Ion Guide mobility measurements through direct infusion (DI). The library consists of CCS and Mass Spectral (MS) data in negative and positive ElectroSpray Ionisation (ESI) mode for 463 and 479 endogenous metabolites, respectively. For both ionisation modes combined, (CCSN2)-C-TW data were obtained for 542 non-redundant metabolites. These data were acquired on two different ion mobility enabled orthogonal acceleration QToF MS systems in two different laboratories, with the majority of the resulting (CCSN2)-C-TW values (from detected compounds) found to be within 1% of one another. Validation of these results against two independent, external (CCSN2)-C-TW data sources and predicted (CSN2)-C-TW values indicated to be within 1-2% of these other values. The same metabolites were then analysed using a rapid reversed-phase ultra (high) performance liquid chromatographic (U(H)PLC) separation combined with IM and MS (IM-MS) thus providing retention time (t(r)), m/z and (CCSN2)-C-TW values (with the latter compared with the DI-IM-MS data). Analytes for which (CCSN2)-C-TW values were obtained by U(H)PLC-IM-MS showed good agreement with the results obtained from DI-IM-MS. The repeatability of the (CCSN2)-C-TW values obtained for these metabolites on the different ion mobility QToF systems, using either DI or LC, encouraged the further evaluation of the U(H)PLC-IM-MS approach via the analysis of samples of rat urine, from control and methotrexatetreated animals, in order to assess the potential of the approach for metabolite identification and profiling in metabolic phenotyping studies. Based on the database derived from the standards 63 metabolites were identified in rat urine, using positive ESI, based on the combination of t(r), (CCSN2)-C-TW and MS data. (C) 2019 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available