4.2 Article

Operation of X-ray gas monitors at the European XFEL

Journal

JOURNAL OF SYNCHROTRON RADIATION
Volume 26, Issue -, Pages 1045-1051

Publisher

INT UNION CRYSTALLOGRAPHY
DOI: 10.1107/S1600577519003795

Keywords

free-electron lasers; photon diagnostics; hard X-rays

Ask authors/readers for more resources

X-ray gas monitors (XGMs) are operated at the European XFEL for non-invasive single-shot pulse energy measurements and average beam position monitoring. They are used for tuning and maintaining the self-amplified spontaneous emission (SASE) operation and for sorting single-shot experimental data according to the pulse-resolved energy monitor data. The XGMs were developed at DESY based on the specific requirements for the European XFEL. In total, six XGM units are continuously in operation. Here, the main principle and experimental setup of an XGM are summarized, and the locations of the six XGMs at the facility are shown. Pulse energy measurements at 0.134nm wavelength are presented, exceeding 1mJ obtained with an absolute measurement uncertainty of 7-10%; correlations between different XGMs are shown, from which a SASE1 beamline transmission of 97% is deduced. Additionally, simultaneous position measurements close to the undulator and at the end of the tunnel are shown, along with the correlation of beam position data simultaneously acquired by an XGM and an imager.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available