4.6 Article

A Non-Invasive Continuous Blood Pressure Estimation Approach Based on Machine Learning

Journal

SENSORS
Volume 19, Issue 11, Pages -

Publisher

MDPI
DOI: 10.3390/s19112585

Keywords

pulse waveform; pulse transit time (PTT); genetic algorithm (GA); machine learning; mean influence value (MIV)

Funding

  1. National Natural Science Found of China [81371713, 8197070720]
  2. Fundamental Research Funds for Central Universities [106112015CDJZR235522]

Ask authors/readers for more resources

Considering the existing issues of traditional blood pressure (BP) measurement methods and non-invasive continuous BP measurement techniques, this study aims to establish the systolic BP and diastolic BP estimation models based on machine learning using pulse transit time and characteristics of pulse waveform. In the process of model construction, the mean impact value method was introduced to investigate the impact of each feature on the models and the genetic algorithm was introduced to implement parameter optimization. The experimental results showed that the proposed models could effectively describe the nonlinear relationship between the features and BP and had higher accuracy than the traditional methods with the error of 3.27 +/- 5.52 mmHg for systolic BP and 1.16 +/- 1.97 mmHg for diastolic BP. Moreover, the estimation errors met the requirements of the Advancement of Medical Instrumentation and British Hypertension Society criteria. In conclusion, this study was helpful in promoting the practical application of methods for non-invasive continuous BP estimation models.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available