4.6 Article

Analysis of a calibration method for non-stationary CVD multi-layered graphene-based gas sensors

Journal

NANOTECHNOLOGY
Volume 30, Issue 38, Pages -

Publisher

IOP Publishing Ltd
DOI: 10.1088/1361-6528/ab2aac

Keywords

graphene; gas sensors; signal steady state and recovery; time-differential signal output; calibration; NO2

Ask authors/readers for more resources

Limitations such as lack of detected stationary signal and slow signal recovery after detection currently affect graphene-based chemi-sensors operating at room temperature. In this work, we model the behavior of a sensor in a test chamber having limited volume and simulating the environmental conditions. From this model, we mathematically derive the calibration method for the sensor. The approach, focused on the time differential of the signal output, is tested on multilayered graphene (MLG)-based sensors towards the chosen target gas (nitrogen dioxide) in the range from 0.12 to 1.32 ppm. MLG acting as sensing layer is synthesized by chemical vapor deposition. Our study paves the route for a wider applicability of the analysis to calibrate the class of devices affected by non-stationary and recovery issues.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available