Journal
MICROMACHINES
Volume 10, Issue 4, Pages -Publisher
MDPI
DOI: 10.3390/mi10040228
Keywords
photoionization detector; microfluidics; microfabrication; volatile organic compound (VOC) detection; toluene
Categories
Funding
- European Community H2020 Framework [643095]
Ask authors/readers for more resources
Government regulations and environmental conditions are pushing the development of improved miniaturized gas analyzers for volatile organic compounds. One of the many detectors used for gas analysis is the photoionization detector (PID). This paper presents the design and characterization of a microfluidic photoionization detector (or mu PID) fabricated using micro milling and electrical discharge machining techniques. This device has no glue and facilitates easy replacement of components. Two materials and fabrication techniques are proposed to produce a layer on the electrodes to protect from ultraviolet (UV) light and possible signal noise generation. Three different microchannels are tested experimentally and their results are compared. The channel with highest electrode area (31.17 mm(2)) and higher volume (6.47 mu L) produces the highest raw signal and the corresponding estimated detection limit is 0.6 ppm for toluene without any amplification unit.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended