4.4 Article

The GLUEX Start Counter Detector

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.nima.2019.02.029

Keywords

GLUEX; Multi-pixel photon counter; Plastic scintillator; Silicon photomultiplier

Funding

  1. Department of Energy [DE-FG02-99ER41065, DE-SC00-13620]
  2. U.S. Department of Energy, Office of Science, Office of Nuclear Physics [DE-AC05-06OR23177]

Ask authors/readers for more resources

The design, simulation, fabrication, calibration, and performance of the GLUEX Start Counter detector are described. The Start Counter was designed to operate at integrated rates of up to 9 MHz with a timing resolution in the range of 500 to 825 ps (FWHM). The Start Counter provides excellent solid angle coverage, a high degree of segmentation for background rejection, and can be utilized in the level 1 trigger for the experiment. It consists of a cylindrical array of 30 thin scintillators with pointed ends that bend towards the beam line at the downstream end. Magnetic field insensitive silicon photomultiplier detectors were used as the light sensors.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available