Measurement of an Analyte Concentration in Test Solution by Using Helmholtz Resonator for Biosensor Applications

Journal

SENSORS
Volume 19, Issue 5, Pages -

Publisher

MDPI
DOI: 10.3390/s19051127

Keywords

analyte concentration; indirect measurement; Helmholtz resonator; resonant frequency; glucose sensor

Funding

  1. BK21 plus program of the Ministry of Education, Korea
  2. Research Project for Railway Technology of the Korea Agency for Infrastructure Technology Advancement (KAIA)

Ask authors/readers for more resources

In this paper, an indirect method of measuring an analyte concentration in a test solution using the resonant frequency change of a Helmholtz resonator is proposed, using a novel architecture of Helmholtz resonator filled with two kinds of fluids (fixed fluid and test solution). Since the analyte concentration yields changes of density and sound speed of the test solution, the resonant frequency of the proposed Helmholtz resonator is affected by the analyte concentration of the test solution. From this effect, the analyte concentration of the test solution can be measured by the spectrum of acoustic resonance of the Helmholtz resonator. The experiment was done using a 3D-printed Helmholtz resonator system with an acoustic power source and detectors, which is consistent with analytical results and showed that the analyte concentration can be measured with higher sensitivity compared to conventional cantilever-type sensors. As an example application, the possibility of measuring glucose concentration of human blood was demonstrated, showing higher sensitivity and relatively low frequency range compared to previous resonance based methods.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available