4.5 Article

The performance of rigid scopes for tracheal intubation: a randomised, controlled trial in patients with a simulated difficult airway

Journal

ANAESTHESIA
Volume 71, Issue 12, Pages 1456-1463

Publisher

WILEY
DOI: 10.1111/anae.13626

Keywords

cervical fracture: intubation techniques; difficult airway algorithm; failed intubation: treatment; rigid fibreoptic scopes; simulated difficult airway

Categories

Funding

  1. Department of Anaesthesiology and Pain Medicine, Inselspital, Bern University Hospital, Switzerland

Ask authors/readers for more resources

We compared the Bonfils (TM) and SensaScope (TM) rigid fibreoptic scopes in 200 patients with a simulated difficult airway randomised to one of the two devices. A cervical collar inhibited neck movement and reduced mouth opening to a mean (SD) of 23 (3) mm. The primary outcome parameter was overall success of tracheal intubation; secondary outcomes included first-attempt success, intubation times, difficulty of intubation, fibreoptic view and side-effects. The mean (95% CI) overall success rate was 88 (80-94)% for the Bonfils and 89 (81-94)% for the SensaScope (p = 0.83). First-attempt intubation success rates were 63 (53-72)% for the Bonfils and 72 (62-81)% for the SensaScope (p = 0.17). Median (IQR [ range]) intubation time was significantly shorter with the SensaScope (34 (20-84 [ 5-240]) s vs. 45 (25-134 [ 12-230]) s), and fibreoptic view was significantly better with the SensaScope (full view of the glottis in 79% with the SensaScope vs. 61% with the Bonfils). This might be explained by its steerable tip and the S-formed shape, contributing to better manoeuvrability. There were no differences in the difficulty of intubation or side-effects.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available