4.5 Article

A false positive case of high-sensitivity cardiac troponin in a patient with acute chest pain: Analytical study of the interference

Journal

CLINICAL BIOCHEMISTRY
Volume 66, Issue -, Pages 103-105

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.clinbiochem.2019.02.002

Keywords

Interference; Cardiac troponin; Immunoassay; Heterophile antibodies

Ask authors/readers for more resources

We report a case of a heterophile antibodies interference in a new high-sensitivity troponin commercial immunoassay (cTNIH Siemens), observed in a patient with possible acute coronary syndrome (ACS). The analytical interference was investigated with standard laboratories procedures. The false positive result was found with different troponin methods and kits. We also investigated the protein sequence of cTnl and no sequence variants were detected. The discordance between clinical pictures and high concentration of cTnl, together with the collaboration between clinicians and laboratory staff avoided possible erroneous diagnosis and further invasive investigations to the patient.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available