Pre-procedural fit-testing of TAVR valves using parametric modeling and 3D printing

Journal

JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY
Volume 13, Issue 1, Pages 21-30

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jcct.2018.09.007

Keywords

Aortic stenosis; 3D printing; Additive manufacturing; Aortic valve; Multi-material printing; Aortic leaflets; Parametric modeling; 3-D printing; Calcifications

Funding

  1. Human Frontier Science Program [HFSP RGY0067]

Ask authors/readers for more resources

Background: Successful transcatheter aortic valve replacement (TAVR) requires an understanding of how a prosthetic valve will interact with a patients anatomy in advance of surgical deployment. To improve this understanding, we developed a benchtop workflow that allows for testing of physical interactions between prosthetic valves and patient-specific aortic root anatomy, including calcified leaflets, prior to actual prosthetic valve placement. Methods: This was a retrospective study of 30 patients who underwent TAVR at a single high volume center. By design, the dataset contained 15 patients with a successful annular seal (defined by an absence of paravalvular leaks) and 15 patients with a sub-optimal seal (presence of paravalvular leaks) on post-procedure transthoracic echocardiogram (TTE). Patients received either a balloon-expandable (Edwards Sapien or Sapien XT, n = 15), or a self-expanding (Medtronic CoreValve or Core Evolut, n = 14, St. Jude Portico, n = 1) valve. Pre-procedural computed tomography (CT) angiograms, parametric geometry modeling, and multi-material 3D printing were utilized to create flexible aortic root physical models, including displaceable calcified valve leaflets. A 3D printed adjustable sizing device was then positioned in the aortic root models and sequentially opened to larger valve sizes, progressively flattening the calcified leaflets against the aortic wall. Optimal valve size and fit were determined by visual inspection and quantitative pressure mapping of interactions between the sizer and models. Results: Benchtop-predicted best fit valve size showed a statistically significant correlation with gold standard CT measurements of the average annulus diameter (n = 30, p < 0.0001 Wilcoxon matched-pairs signed rank test). Adequateness of seal (presence or absence of paravalvular leak) was correctly predicted in 11/15 (73.3%) patients who received a balloon-expandable valve, and in 9/15 (60%) patients who received a self-expanding valve. Pressure testing provided a physical map of areas with an inadequate seal; these corresponded to areas of paravalvular leak documented by post-procedural transthoracic echocardiography. Conclusion: We present and demonstrate the potential of a workflow for determining optimal prosthetic valve size that accounts for aortic annular dimensions as well as the active displacement of calcified valve leaflets during prosthetic valve deployment. The workflow's open source framework offers a platform for providing predictive insights into the design and testing of future prosthetic valves.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available