4.5 Article Proceedings Paper

Uncertainty analysis of response functions and γ-backgrounds on Tion and t0 measurements from Cherenkov neutron detectors at the National Ignition Facility (NIF)

Journal

REVIEW OF SCIENTIFIC INSTRUMENTS
Volume 89, Issue 10, Pages -

Publisher

AMER INST PHYSICS
DOI: 10.1063/1.5038816

Keywords

-

Funding

  1. U.S. Department of Energy by Lawrence Livermore National Laboratory [DE-AC52-07NA27344]
  2. U.K. Ministry of Defence

Ask authors/readers for more resources

Cherenkov radiators deployed to measure the neutron time-of-flight spectrum have response times associated with the neutron transit across the detector and are free from long time response tails characteristic of scintillation detectors. The Cherenkov radiation results from simple physical processes which makes them amenable to high fidelity Monte Carlo simulation. The instrument response function of neutron time-of-flight systems is a major contributor to both the systematic and statistical uncertainties of the parameters used to describe these spectra; in particular, the first and second moments of these distributions are associated with arrival time, t(0), and ion temperature, T-ion. We present the results of uncertainty analysis showing the significant reduction of the uncertainty in determining these quantities in the Cherenkov detector system recently deployed at NIF. The increased sensitivity to gamma radiation requires additional consideration of the effect of this background to the uncertainties in both t(0) and T-ion. Published by AIP Publishing.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available