4.1 Article

Insertion of peripherally inserted central catheters with intracavitary electrocardiogram guidance: A randomized multicenter study in China

Journal

JOURNAL OF VASCULAR ACCESS
Volume 20, Issue 5, Pages 524-529

Publisher

SAGE PUBLICATIONS LTD
DOI: 10.1177/1129729818819732

Keywords

Peripherally inserted central catheters; intracavitary electrocardiogram; tip location; ultrasound guidance

Ask authors/readers for more resources

Introduction: Ultrasound-guided venipuncture and tip location by intracavitary electrocardiogram have many advantages during the insertion of peripherally inserted central catheters, both in terms of safety and cost-effectiveness. Recently, a new tip-conductive peripherally inserted central catheters and new Doppler ultrasound device integrated with intracavitary electrocardiogram have been introduced into clinical practice in China. A randomized multicenter study (clinical trial no. NCT03230357) was performed to verify the feasibility and accuracy of intracavitary electrocardiogram, as performed with this new peripherally inserted central catheters and device. Methods: Our study enrolled a total of 2250 adult patients in 10 different Chinese hospitals. The patients were randomly assigned to either the study group (intracavitary electrocardiogram) or the control group (anatomical landmark guidance) in a 2:1 allocation. Ultrasound was used in both groups for venipuncture and tip navigation. All patients underwent chest X-ray after the procedure to verify the position of the catheter tip. Results: No insertion-related complications were reported in either group. In the study group, first-attempt successful tip location was 91.7% (95% confidence interval: 90.3%-93.1%), significantly higher than 78.9% (95% confidence interval: 76.0%-81.9%) observed in the control group (p < 0.001). As evaluated by post-procedural chest X-ray, tip location in the study group had a sensitivity of 99.3% (95% confidence interval: 98.8%-99.7%), significantly higher than 86.8% (95% confidence interval: 84.4%-89.2%) observed in the anatomical landmark group (p < 0.001). Conclusion: These results indicated that during peripherally inserted central catheters insertion in adult patients, tip location with intracavitary electrocardiogram guidance, as carried out by a new tip-conductive peripherally inserted central catheters and intracavitary electrocardiogram integrated ultrasound device, was more effective and more accurate than tip location using anatomical landmarks.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.1
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available