4.3 Article

Comparison of macronutrient content in human milk measured by mid-infrared human milk analyzer and reference methods

Journal

JOURNAL OF PERINATOLOGY
Volume 39, Issue 3, Pages 497-503

Publisher

NATURE PUBLISHING GROUP
DOI: 10.1038/s41372-018-0291-8

Keywords

-

Funding

  1. Nestle Research

Ask authors/readers for more resources

Objective The study aims at evaluating mid-infrared human milk analyzer (HMA) accuracy and precision, in human milk (HM). Study design Rose-Gottlieb, high-performance anion exchange chromatography-pulsed amperometric detection (HPAEC-PAD), Kjeldahl and amino acid analysis (AA) were selected as references for total fat, lactose and total protein determination. Results No significant difference was observed in lactose content between HMA and HPAEC-PAD. Significant differences were observed in fat and protein content between HMA and reference methods. However, the difference in fat content was lower than 12%, and therefore within the variability declared by supplier. For protein determination, the BCA protein assay was selected. No significant differences were observed in total protein content measured by BCA assay, Kjeldahl and AA methods. Conclusions HMA was reliable for the quantification of total fat and lactose content, but not for total protein one. The latter was measured by BCA assay, which yielded comparable results to Kjeldahl and AA methods.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available