Coronary artery calcium scoring with photon-counting CT: first in vivo human experience

Journal

INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING
Volume 35, Issue 4, Pages 733-739

Publisher

SPRINGER
DOI: 10.1007/s10554-018-1499-6

Keywords

Coronary artery calcium score; Photon-counting CT; Radiation dose reduction

Funding

  1. NIH intramural research program [ZIACL090019, ZIAEB000072]
  2. Siemens Healthcare GmbH (Forchheim, Germany)

Ask authors/readers for more resources

To evaluate the performance of photon-counting detector (PCD) computed tomography (CT) for coronary artery calcium (CAC) score imaging at standard and reduced radiation doses compared to conventional energy-integrating detector (EID) CT. A dedicated cardiac CT phantom, ten ex vivo human hearts, and ten asymptomatic volunteers underwent matched EID and PCD CT scans at different dose settings without ECG gating. CAC score, contrast, and contrast-to-noise ratio (CNR) were calculated in the cardiac CT phantom. CAC score accuracy and reproducibility was assessed in the ex vivo hearts. Standard radiation dose (120kVp, reference mAs=80) in vivo CAC scans were compared against dose-reduced CAC scans (75% dose reduction; reference mAs=20) for image quality and CAC score reproducibility. Interstudy agreement was assessed by using intraclass correlation (ICC), linear regression, and Bland-Altman analysis with 95% confidence interval limits of agreement (LOA). Calcium-soft tissue contrast and CNR were significantly higher for the PCD CAC scans in the cardiac CT phantom (all P<0.01). Ex vivo hearts: CAC score reproducibility was significantly higher for the PCD scans at the lowest dose setting (50mAs) (P=0.002); score accuracy was similar for both detector systems at all dose settings. In vivo scans: the agreement between standard dose and low dose CAC score was significantly better for the PCD than for the EID with narrower LOA in Bland-Altman analysis, linear regression slopes closer to 1 (0.96 vs. 0.84), and higher ICC values (0.98 vs. 0.93, respectively). Phantom and in vivo human studies showed PCD may significantly improve CAC score image quality and/or reduce CAC score radiation dose while maintaining diagnostic image quality.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available