How accurate and precise are CT based measurements of iodine concentration? A comparison of the minimum detectable concentration difference among single source and dual source dual energy CT in a phantom study

Journal

EUROPEAN RADIOLOGY
Volume 29, Issue 4, Pages 2069-2078

Publisher

SPRINGER
DOI: 10.1007/s00330-018-5736-0

Keywords

Multidetector computed tomography; Iodine; Phantom imaging

Ask authors/readers for more resources

ObjectivesTo assess the impact of scan- and patient-related factors on the error and the minimum detectable difference in iodine concentration among different generations of single-source (SS) fast kV-switching and dual-source (DS) dual-energy CT (DECT).MethodsLesions having eight different iodine concentrations (0.2-4 mgI/mL) were emulated in a 3D-printed phantom of medium and large size. Each combination of concentration and size was scanned in dual-energy mode on four different SS and DS DECTs. Radiation doses were 7 and 10 mGy (medium size) and 10, 13, and 16 mGy (large size). Iodine maps were reconstructed with filtered back projection (FBP) and vendor-specific iterative reconstruction algorithms (IRs). Absolute error of iodine quantification (E) was measured. Multivariate regression models determined the influence of CT scanner, iodine concentration, phantom size, radiation dose, and reconstruction algorithm on E. The minimum detectable difference in iodine concentration (ICmin) under the same imaging conditions (intra-conditional) and among different imaging conditions (inter-conditional) was calculated.ResultsThe error was significantly lower in current than in previous DECT generations (p < 0.001). For all CT scanner conditions, the error was significantly higher with increasing phantom size and decreasing radiation dose (p < 0.001). Iodine concentration only significantly affected the error for SS DECT (p < 0.001). ICmin depended on patient- and scan-related factors and ranged from 0.4 to 1.5 mgI/mL.ConclusionsPatient- and scan-related factors have a significant impact on the error and minimum detectable difference in iodine concentration within and among SS fast kV-switching and DS DECT.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available