4.0 Article

Double-Focusing Thermal Triple-Axis Spectrometer at the NCNR

Publisher

NATL INST STANDARDS & TECHNOLOGY-NIST
DOI: 10.6028/jres.117.002

Keywords

Inelastic neutron scattering; neutron diffraction; polarized neutrons; position sensitive detector; thermal neutron triple-axis spectrometer

Ask authors/readers for more resources

The new thermal triple-axis spectrometer at the NIST Center for Neutron Research (NCNR) is located at the BT-7 beam port. The 165 mm diameter reactor beam is equipped with a selection of Soller collimators, beam-limiters, and a pyrolytic graphite (PG) filter to tailor the beam for the dual 20x20 cm(2) double-focusing monochromator system that provides monochromatic fluxes exceeding 10(8) n/cm(2)/s onto the sample. The two monochromators installed are PG(002) and Cu(220), which provide incident energies from 5 meV to above 500 meV. The computer controlled analyzer system offers six standard modes of operation, including a diffraction detector, a position-sensitive detector (PSD) in diffraction mode, horizontal energy focusing analyzer with detector, a Q-E mode employing a flat analyzer and PSD, a constant-E mode with the analyzer crystal system and PSD, and a conventional mode with a selection of Soller collimators and detector. Additional configurations for specific measurement needs are also available. This paper discusses the capabilities and performance for this new state-of-the-art neutron spectrometer.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.0
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available