Performance of the Advia Centaur second-generation troponin assay TnI-Ultra compared with the first-generation cTnI assay

Journal

ANNALS OF CLINICAL BIOCHEMISTRY
Volume 45, Issue -, Pages 316-317

Publisher

ROYAL SOC MEDICINE PRESS LTD
DOI: 10.1258/acb.2007.007209

Keywords

-

Ask authors/readers for more resources

Background: A cardiac troponin concentration above the 99th percentile limit of a reference population is a sensitive marker of myocardial necrosis. Current guidelines require troponin assays to have a total imprecision of <= 10% at the 99th percentile limit. In this study, the Advia Centaur second-generation TnI-Ultra assay was validated and compared with its predecessor the cardiac troponin I (cTnI) assay, with a focus on the current guidelines for diagnosis of acute myocardial damage. Methods: An imprecision profile of the TnI-Ultra assay was evaluated by analysing different pools over 20 days. The imprecision of the cTnI assay was used as comparison. The reference range was established based on Tnl-Ultra analysis in 221 individuals. Results: The cTnI concentration that could be determined with a total imprecision of 10% was 0.05 mu g/L for the Tnl-Ultra assay and 0.3 mu g/L for the cTnI assay. The 99th percentile limit in the distribution of a reference population was 0.06 mu g/L as determined with the Tnl-Ultra assay. Conclusions: The Tnl-Ultra assay provides significantly improved sensitivity when compared with the cTnI assay and a total imprecision of <= 10% is obtained at the 99th percentile limit of value distribution of a reference population. Using the TnI-Ultra assay, slightly increased cTnI concentration can be detected reliably following the current guidelines.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available