Journal
JOURNAL OF ADVANCED PROSTHODONTICS
Volume 6, Issue 1, Pages 35-38Publisher
KOREAN ACAD PROSTHODONTICS
DOI: 10.4047/jap.2014.6.1.35
Keywords
Microleakage; Implant-abutment connection; Screw loosening
Categories
Ask authors/readers for more resources
PURPOSE. This study aimed to evaluate the correlation between microleakage and screw loosening at different types of implant-abutment connections and/or geometries measuring the torque values before and after the leakage tests. MATERIALS AND METHODS. Three different abutment types (Intenal hex titanium, internal hex zirconium, morse tapered titanium) with different geometries were connected to its own implant fixture. All the abutments were tightened with a standard torque value then the composition was connected to the modified fluid filtration system. After the measurements of leakage removal torque values were re-measured. Kruskal-wallis test was performed for non-parametric and one-way ANOVA was performed for parametric data. The correlation was evaluated using Spearman Correlation Test (alpha=0.05). RESULTS. Significantly higher microleakage was found at the connection of implant-internal hex zirconium abutment. Observed mean torque value loss was also significantly higher than other connection geometries. Spearman tests revealed a significant correlation between microleakage and screw loosening. CONCLUSION. Microleakage may provoke screw loosening. Removing torque values rationally decrease with the increase of microleakage.
Authors
I am an author on this paper
Click your name to claim this paper and add it to your profile.
Reviews
Recommended
No Data Available