4.6 Article

Computer-Aided Bleeding Detection in WCE Video

Journal

IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS
Volume 18, Issue 2, Pages 636-642

Publisher

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/JBHI.2013.2257819

Keywords

Bleeding detection; superpixel; wireless capsule endoscopy (WCE)

Funding

  1. National Science Foundation of China (NSFC) [61203253]

Ask authors/readers for more resources

Wireless capsule endoscopy (WCE) can directly take digital images in the gastrointestinal tract of a patient. It has opened a new chapter in small intestine examination. However, a major problem associated with this technology is that too many images need to be manually examined by clinicians. Currently, there is no standard for capsule endoscopy image interpretation and classification. Most state-of-the-art CAD methods often suffer from poor performance, high computational cost, or multiple empirical thresholds. In this paper, a new method for rapid bleeding detection in theWCE video is proposed. We group pixels through superpixel segmentation to reduce the computational complexity while maintaining high diagnostic accuracy. Feature of each superpixel is extracted using the red ratio in RGB space and fed into support vector machine for classification. Also, the influence of edge pixels has been removed in this paper. Comparative experiments show that our algorithm is superior to the existing methods in terms of sensitivity, specificity, and accuracy.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available