4.7 Article

Ultra-fast cell counters based on microtubular waveguides

Journal

Scientific Reports
Volume 7, Issue -, Pages -

Publisher

NATURE PUBLISHING GROUP
DOI: 10.1038/srep41584

Keywords

-

Funding

  1. Hamburg Landesforschungsforderung via the project ReAd! Me

Ask authors/readers for more resources

We present a radio-frequency impedance-based biosensor embedded inside a semiconductor microtube for the in-flow detection of single cells. An impedance-matched tank circuit and a tight wrapping of the electrodes around the sensing region, which creates a close, leakage current-free contact between cells and electrodes, yields a high signal-to-noise ratio. We experimentally show a twofold improved sensitivity of our three-dimensional electrode structure to conventional planar electrodes and support these findings by finite element simulations. Finally, we report on the differentiation of polystyrene beads, primary mouse T lymphocytes and Jurkat T lymphocytes using our device.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available