Amine-intercalated montmorillonite matrices for enzyme immobilization and biosensing applications

Journal

RSC ADVANCES
Volume 2, Issue 5, Pages 2112-2118

Publisher

ROYAL SOC CHEMISTRY
DOI: 10.1039/c2ra01225a

Keywords

-

Ask authors/readers for more resources

Clay based biosensors were developed using montmorillonite (Mont) modified with methyl (M) and dimethylamine (DM). X-ray diffraction, Fourier transform infrared spectroscopy, zeta potential and thermal gravimetric measurements were used to characterize the modified clays. After immobilization of glucose oxidase (GOx) via clay on the glassy carbon electrode, its application as a glucose biosensor was investigated in detail. The best response characteristics were obtained by DM-Mont and optimization of enzyme amount, reproducibility of biosensor fabrication, repeatability of measurements and operational stability were all evaluated. The optimized biosensor showed a very good linearity between 0.05 mM and 1.0 mM, a 7 s response time and a limit of detection to glucose of 0.038 mM. Also, kinetic parameters and stabilities were determined. Apparent K-m and I-max values were found as 0.73 mM and 2.955 mu A, respectively. As well as batch configuration, the DM-Mont/GOx biosensor was successfully applied in the flow injection analysis mode. Finally, the performance of the DM-Mont/GOx biosensor to analyze glucose in a wine sample was compared with HPLC.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available