4.5 Article

Edwards SAPIEN 3 valve

Journal

EUROINTERVENTION
Volume 8, Issue -, Pages Q83-Q87

Publisher

EUROPA EDITION
DOI: 10.4244/EIJV8SQA15

Keywords

-

Ask authors/readers for more resources

Building on the established success with the SAPIEN, SAPIEN XT and earlier prototypic transcatheter heart valves (THV) the newest balloon-expandable valve incorporates a number of new and enhanced features intended to reduce the risk of vascular injury, to reduce paravalvular regurgitation, and to facilitate rapid and accurate positioning and implantation. The SAPIEN 3 THV incorporates a cobalt chromium stent, bovine pericardial leaflets, and both an inner and new outer polyethylene terephthalate sealing cuff. The delivery system incorporates an active three-dimensional coaxial positioning catheter, and is compatible with a 14 Fr expandable sheath.