4.3 Review

How to Perform and Interpret a High-resolution Anorectal Manometry Test

Journal

JOURNAL OF NEUROGASTROENTEROLOGY AND MOTILITY
Volume 22, Issue 1, Pages 46-59

Publisher

KOREAN SOC NEUROGASTROENTEROLOGY & MOTILITY
DOI: 10.5056/jnm15168

Keywords

Anal canal; Constipation; Fecal incontinence; Manometry; Physiology

Funding

  1. NIH from the National Institutes of Health (NIH) [R01 DKDK78924]
  2. Soonchunhyang University Research Fund

Ask authors/readers for more resources

High-resolution anorectal manometry (HR-ARM) and high-definition anorectal manometry (HD-ARM) catheters have closely spaced water-perfused or solid state circumferentially-oriented pressure sensors that provide much better spatiotemporal pressurization than non-high resolution catheters. This is a comprehensive review of HR-ARM and HD-ARM anorectal manometry catheter systems, the methods for conducting, analyzing, and interpreting HR-ARM and HD-ARM, and a comparison of HR-ARM with non-high resolution anorectal manometry. Compared to non-high resolution techniques, HR-ARM and HD-ARM studies take less time and are easier to interpret. However, HR-ARM and HD-ARM catheters are more expensive and fragile and have a shorter lifespan. Further studies are needed to refine our understanding of normal values and to rigorously evaluate the incremental clinical utility of HR-ARM or HD-ARM compared to non-high resolution manometry.