4.6 Article

Newly designed J-shaped tip guidewire: A preliminary feasibility study in wire-guided cannulation

Journal

WORLD JOURNAL OF GASTROENTEROLOGY
Volume 19, Issue 28, Pages 4531-4536

Publisher

BAISHIDENG PUBL GRP CO LTD
DOI: 10.3748/wjg.v19.i28.4531

Keywords

J-shaped tip guidewire; Wire-guided cannulation; Endoscopic retrograde cholangiopancreatography; Biliary tract; Cannulation technique; Perforation

Funding

  1. Toho University Ohashi Medical Center

Ask authors/readers for more resources

AIM: To perform wire-guided cannulation using a newly designed J-shaped tip guidewire, and to verify feasibility and safety for use. METHODS: The study was conducted on endoscopic retrograde cholangiopancreatography (ERCP) patients with naive papilla undergoing diagnosis and treatment of biliary diseases between September 2011 and July 2012. We performed ERCP in a succession of 50 cases with a J-shaped tip guidewire. The first insertion attempt began with a trainee who had 5 min to complete cannulation, followed if necessary by the trainer for another 5 min. We assessed the primary success rate of selective biliary cannulation within 10 min and adverse events such as post-ERCP pancreatitis (PEP), bleeding or perforation. RESULTS: The primary success rate was 90% (45/50) within 10 min, the initial success rate within 5 min by trainee staff was 76% (38/50). The rate of PEP was 6% (3/50), but all 3 cases were mild pancreatitis. All patients were managed successfully with conservative treatment. There was no bleeding or perforation. CONCLUSION: A newly designed J-shaped tip guidewire has the possibility to facilitate selective biliary cannulation for ERCP and appears to be safe. (C) 2013 Baishideng. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available