Calibration and performance of a fast-scanning DSC-Project RHC

Journal

THERMOCHIMICA ACTA
Volume 530, Issue -, Pages 64-72

Publisher

ELSEVIER
DOI: 10.1016/j.tca.2011.12.001

Keywords

Fast-scanning calorimetry; Thermal lag; Zero-heating rate temperature calibration; Heat flow rate calibration; Liquid crystal

Funding

  1. Fund for Scientific Research-Flanders, Belgium (F.W.O. Vlaanderen)

Ask authors/readers for more resources

This paper presents the calibration and evaluation of a Rapid Heat-Cool DSC (RHC), a fast-scanning calorimeter capable of controlled heating rates of up to 2000 K min(-1). As the use of high scan rates increases the influence of thermal lag, a thorough analysis of the instrument performance was made. This encompasses a rigorous temperature calibration procedure, a study of the influence of sample mass and heating rate on the extrapolated onset temperature, and the analysis of the symmetry of the thermal lag in heating and cooling using phase transitions in liquid crystals. The calibration of the heat flow rate is evaluated using indium for a broad range of sample masses and heating rates, and verified by heat capacity measurements on polystyrene for a wide temperature range. The performance of the RHC proves to be very good, with estimated errors on temperature and heat flow rate (heat capacity) of 0.1 K and 2%, respectively. (C) 2011 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available