4.5 Article

Calibration of differential scanning calorimeters: A comparison between indium and diphenylacetic acid

Journal

THERMOCHIMICA ACTA
Volume 497, Issue 1-2, Pages 72-76

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.tca.2009.08.010

Keywords

DSC; Calibration; Indium; Diphenylacetic acid

Ask authors/readers for more resources

The close proximity in melting temperature of the LGC Limited DSC standards indium and diphenylacetic acid, has enabled a direct assessment to be made of any differences resulting from the use of a metal or an organic compound in the calibration of DSC equipment. Following calibration with indium, the equilibrium fusion temperatures for diphenylacetic acid, were determined by both the stepwise heating and extrapolation to zero heating rate methods. The results were in excellent agreement with the certificate values and established that indium may be used as a calibrant when making accurate DSC measurements on Organic materials in the same temperature range and that it has the advantage that it is non-volatile and can be used a number of times without significant change. Similar agreement was obtained in the measurement of the enthalpy of fusion, although the larger heat capacity change on fusion of diphenylacetic acid resulted in a greater uncertainty than with indium. (C) 2009 Elsevier B.V. All rights reserved.