4.7 Article

A centrifugal microanalysis system for the determination of nitrite and hexavalent chromium

Journal

TALANTA
Volume 80, Issue 2, Pages 670-675

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.talanta.2009.07.046

Keywords

Microanalysis; Centrifugal microfluidics; Nitrite; Hexavalent chromium; Environmental analysis

Funding

  1. Natural Sciences and Engineering Research Council

Ask authors/readers for more resources

A centrifugal microanalysis system using spectrophotometric detection has been developed for the determination of nitrite and chromium VI. The system was designed to illustrate the potential of centrifugal microfluidic systems for on-site (field) analysis of waters. The performance was significantly better than manual/visual field test systems and approaches that of bench top systems with detection limits of 0.008 mg L-1 NO2--N and 0.03 mg L-1 Cr6+. The system used pre-loaded solid reagents on the disc, providing ease of use, and could process up to 24 samples simultaneously. Precision was enhanced by the use of a multiwavelength ratiometric blank estimation technique. (C) 2009 Elsevier B.V. All rights reserved.