4.7 Article

Determination of bismuth in solid samples by hydride generation atomic fluorescence spectrometry with a dielectric barrier discharge atomizer

Journal

TALANTA
Volume 80, Issue 1, Pages 139-142

Publisher

ELSEVIER
DOI: 10.1016/j.talanta.2009.06.038

Keywords

Dielectric barrier discharge; Atomizer; Hydride generation; Atomic fluorescence spectrometry

Funding

  1. innovation Method Fund of China [20081M040600]
  2. NSFC [20535020, 20775041]

Ask authors/readers for more resources

An atmospheric pressure dielectric barrier discharge (DBD) atomizer was investigated for bismuth (Bi) determination with hydride generation (HG) atomic fluorescence spectrometry (AFS). The characteristics of the atomizer and the effects of experimental parameters, including observation height, discharge power, flow rate of discharge gas and AFS carrier gas were optimized. The linear range of present method for Bi determination is 0.5-300.0 mu g L-1 with a detection limit of 0.07 mu g L-1 (3 sigma). The method was validated by the analysis of reference materials (GBW08517 and GSB-14) and the results agreed well with the reference values. The established method was applied to the determination of Bi in ore, soil and ash samples. (C) 2009 Elsevier B.V. All rights reserved.