Benefit of replacing the Sigma-60 by the Sigma-Eye applicator A Monte Carlo-based uncertainty analysis

Journal

STRAHLENTHERAPIE UND ONKOLOGIE
Volume 189, Issue 1, Pages 74-80

Publisher

URBAN & VOGEL
DOI: 10.1007/s00066-012-0241-x

Keywords

Applicators; Cervical cancer; Hyperthermia treatment planning; Monte Carlo analysis; Thermal dose

Ask authors/readers for more resources

To investigate the clinical benefit of replacing the BSD-2000 Sigma-60 with the Sigma-Eye applicator, taking into account effects of uncertainties in tissue and water bolus parameters. For 20 patients, specific absorption rate (SAR) and temperature distributions were calculated and optimized, based on computed tomography (CT) scans in treatment position. The impact of uncertainties on predicted distributions was studied using a Monte Carlo uncertainty assessment. Replacing the Sigma-60 by the Sigma-Eye applicator resulted in a higher SAR in the tumor [on average a decrease of the hotspot tumor quotient (HTQ) by 24%; p < 0.001], and higher temperatures (T90: +0.4A degrees C, p < 0.001; T50: +0.6A degrees C, p < 0.001) using literature values and SAR optimization. When temperature optimization (T90) was used, a larger average increase was found (T90: +0.7A degrees C, p < 0.001; T50: +0.8A degrees C, p < 0.001). When taking into account uncertainties, a decrease of 23% in median HTQ (p < 0.001) and an increase in T50 and T90 of 0.4A degrees C (p < 0.001) could be demonstrated. Based on this uncertainty analysis, significant and clinically relevant improvements in HTQ and tumor temperature were achieved when replacing the Sigma-60 by the Sigma-Eye applicator.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available