4.1 Article

The Risk of Hardware Infection in Deep Brain Stimulation Surgery Is Greater at Impulse Generator Replacement than at the Primary Procedure

Journal

STEREOTACTIC AND FUNCTIONAL NEUROSURGERY
Volume 91, Issue 1, Pages 56-65

Publisher

KARGER
DOI: 10.1159/000343202

Keywords

Deep brain stimulation; Infection; Parkinson's disease; Dystonia; Reoperation

Funding

  1. Edmond J. Safra Philanthropic Foundation
  2. Monument Trust
  3. Parkinson's Appeal, UK

Ask authors/readers for more resources

Background/Aims: Infection of implanted hardware after deep brain stimulation (DBS) has a significant impact on patient morbidity. We examined all patients who underwent DBS procedures over the last 9 years in our centre to assess the infection rate and possible factors related to surgery that may predispose to infection. Methods: Surgical reports and clinical notes were reviewed in 273 consecutive patients who underwent a total of 519 DBS-related procedures in our institute between November 2002 and September 2011. Results:Sixteen separate hardware-related infections occurred in 11 patients. Infections occurred in 3% of all procedures and 4% of all patients. The infection rate after implantable pulse generator (IPG) replacement surgery was more than three times higher than after de novo DBS surgery. In addition, male patients were more likely to develop device-related infections. Conclusions: It is unclear why infection rates should be more than three times higher after IPG replacement surgery than after the de novo procedure. The former is a shorter and simpler procedure to conduct. Perhaps the use of better antimicrobial protection and rechargeable batteries may be useful strategies to reduce infections following IPG revision surgery. Copyright (C) 2012 S. Karger AG, Basel

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.1
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available