4.5 Article

Comparison of 3-Dimensional Spinal Reconstruction Accuracy Biplanar Radiographs With EOS Versus Computed Tomography

Journal

SPINE
Volume 37, Issue 16, Pages 1391-1397

Publisher

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/BRS.0b013e3182518a15

Keywords

scoliosis; imaging; EOS; computed tomography (CT); 3D reconstructions; low radiation; biplanar slot-scanner; validation; accuracy

Funding

  1. EOS Imaging, France

Ask authors/readers for more resources

Study Design: Experimental study for systematic evaluation of 3-dimensional (3D) reconstructions from low-dose digital stereoradiography. Objective. To assess the accuracy of EOS (EOS Imaging, Paris, France) 3-dimensional (3D) reconstructions compared with 3D computed tomography (CT) and the effect spine positioning within the EOS unit has on reconstruction accuracy. Summary of Background Data. Scoliosis is a 3D deformity, but 3D morphological analyses are still rare. A new low-dose radiation digital stereoradiography system (EOS) was previously evaluated for intra/interobserver variability, but data are limited for 3D reconstruction accuracy. Methods. Three synthetic scoliotic phantoms (T1-pelvis) were scanned in upright position at 0 degrees, +/- 5 degrees, and +/- 10 degrees of axial rotation within EOS and in supine position using CT. Three-dimensional EOS reconstructions were superimposed on corresponding 3D computed tomographic reconstructions. Shape, position, and orientation accuracy were assessed for each vertebra and the entire spine. Additional routine planer clinical deformity measurements were compared: Cobb angle, kyphosis, lordosis, and pelvic incidence. Results. Mean EOS vertebral body shape accuracy was 1.1 +/- 0.2 mm (maximum 4.7 mm), with 95% confidence interval of 1.7 mm. Different anatomical vertebral regions were modeled well with root-mean-square (RMS) values from 1.2 to 1.6 mm. Position and orientation accuracy of each vertebra were high: RMS offset was 1.2 mm (maximum 3.7 mm) and RMS axial rotation was 1.9 degrees (maximum 5.8 degrees). There was no significant difference in each of the analyzed parameters (P > 0.05) associated with varying the rotational position of the phantoms in EOS machine. Planer measurements accuracy was less than 1 degrees mean difference for pelvic incidence, Cobb angle (mean 1.6 degrees/maximum 3.9 degrees), and sagittal kyphosis (mean less than 1 degrees, maximum 4.9 degrees). Conclusion. The EOS image acquisition and reconstruction software provides accurate 3D spinal representations of scoliotic spinal deformities. The results of this study provide spinal deformity surgeons evidence pertaining to this new upright 3D imaging technology that may aid in the clinical diagnosis and decision making for patients with scoliosis.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available