4.7 Article

A green cellulose paper based glucose amperometric biosensor

Journal

SENSORS AND ACTUATORS B-CHEMICAL
Volume 193, Issue -, Pages 536-541

Publisher

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2013.11.054

Keywords

Cellulose paper based amperometric biosensor; Physical adsorption; Glucose oxidase

Ask authors/readers for more resources

A simple low cost green biosensor configuration comprising of a hydrophilic cellulose paper disk with immobilised glucose oxidase (GOx) via adsorption step, placed on top of a screen printed carbon electrode (SPCE) was developed. This biosensor configuration allowed for low volume of glucose sample (5 mu L) to be analysed. Cellulose paper was also used as the pre-storage reagent matrix for 0.1 M phosphate buffer solution (PBS, pH 7.0) and 10 mM soluble ferrocene monocarboxylic acid mediator. This biosensor exhibited a linear dynamic calibration range of 1 to 5 mM glucose (r(2) = 0.971), with a limit of detection of 0.18 mM and retained 98% of its signal after a period of four months. In addition, its performance was demonstrated in the analysis of selected commercial soda beverages. The glucose concentrations obtained by the biosensor corroborated well with an independent high performance liquid chromatographic (HPLC) method. (C) 2013 Elsevier B.V. All rights reserved.