4.7 Article Proceedings Paper

Multi-analytic grating coupler biosensor for differential binding analysis

Journal

SENSORS AND ACTUATORS B-CHEMICAL
Volume 144, Issue 2, Pages 413-417

Publisher

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2009.02.003

Keywords

Optical grating coupler; Multi-channel biorecognition; On-chip reference

Ask authors/readers for more resources

In this paper, a multiple-channel extension of a dual-grating Coupler biosensor is presented as a Solution for the problem of resolving different selectivities, Usual when heterogeneous samples are analyzed. Several differently functionalized channels can perform quantitative analysis of competiting recognition events, Suppress shifts due to buffer changes and even monitorize drifts coming from the light Source. Here, the multiple-channel approach is developed and proven for a four-channel configuration, providing a resolution limit of 10(-5) Refractive index Units (RIU) and with an a potentially Unlimited scalability. Finally, a differential HSA recognition event is monitored, at both an IgG functionalized channel and at a blocked one. (C) 2009 Elsevier B.V. All lights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available