4.7 Article

Detection of Infertility-related Neutralizing Antibodies with a Cell-free Microfluidic Method

Journal

SCIENTIFIC REPORTS
Volume 5, Issue -, Pages -

Publisher

NATURE PUBLISHING GROUP
DOI: 10.1038/srep16551

Keywords

-

Funding

  1. European Research Council (ERC) [203428]
  2. Merck Serono

Ask authors/readers for more resources

The unwanted emergence of neutralizing antibodies (nAbs) against an endogenous or a therapeutic protein can result in deficiency diseases or therapy failure. Here, we developed a cell-free microfluidic method for the sensitive detection and quantification of nAbs in human serum that are associated with infertility. We used cell-derived vesicles containing the luteinizing hormone (LH)/choriogonadotropin receptor (LHHCGR) to detect nAbs against LH. The method exploits the entire cellular signal amplification mechanism, and facilitates the detection of as little as 0.44 nM of LH-nAb (K-d 1.5 nM) in human serum matrix within only 15 minutes. In addition, dose-response curves can be generated in less than 2 hours to evaluate the nAB concentration and dissociation constant. The developed system is devoid of problems associated with cell-based assays and we believe that this simple effect-directed analysis can be used in clinical environments, and is adaptable to other hormones or cytokines and their respective nAbs.