Calibration of low-temperature ac susceptometers with a copper cylinder standard

Journal

REVIEW OF SCIENTIFIC INSTRUMENTS
Volume 81, Issue 2, Pages -

Publisher

AMER INST PHYSICS
DOI: 10.1063/1.3309779

Keywords

calibration; copper; eddy currents; magnetic susceptibility; magnetic variables measurement

Funding

  1. Consolider Project [CSD2007-00041, 2009SGR-1292]
  2. ICREA Funding Source: Custom

Ask authors/readers for more resources

A high-quality low-temperature ac susceptometer is calibrated by comparing the measured ac susceptibility of a copper cylinder with its eddy-current ac susceptibility accurately calculated. Different from conventional calibration techniques that compare the measured results with the known property of a standard sample at certain fixed temperature T, field amplitude H-m, and frequency f, to get a magnitude correction factor, here, the electromagnetic properties of the copper cylinder are unknown and are determined during the calibration of the ac susceptometer in the entire T, H-m, and f range. It is shown that the maximum magnitude error and the maximum phase error of the susceptometer are less than 0.7% and 0.3 degrees, respectively, in the region T=5-300 K and f=111-1111 Hz at H-m=800 A/m, after a magnitude correction by a constant factor as done in a conventional calibration. However, the magnitude and phase errors can reach 2% and 4.3 degrees at 10 000 and 11 Hz, respectively. Since the errors are reproducible, a large portion of them may be further corrected after a calibration, the procedure for which is given. Conceptual discussions concerning the error sources, comparison with other calibration methods, and applications of ac susceptibility techniques are presented.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available