Systematic comparison of δ13C measurements of testosterone and derivative steroids in a freeze-dried urine candidate reference material for sports drug testing by gas chromatography/combustion/isotope ratio mass spectrometry and uncertainty evaluation using four different metrological approaches

Journal

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
Volume 25, Issue 11, Pages 1641-1651

Publisher

WILEY-BLACKWELL
DOI: 10.1002/rcm.4983

Keywords

-

Ask authors/readers for more resources

An alternative calibration procedure for use when performing carbon isotope ratio measurements by gas chromatography/combustion/isotope ratio mass spectrometry (GC/C/IRMS) has been developed. This calibration procedure does not rely on the corrections in-built in the instrument software, as the carbon isotope ratios of a sample are calculated from the measured raw peak areas. The method was developed for the certification of a urine reference material for sports drug testing, as the estimation of measurement uncertainty is greatly simplified. To ensure that the method is free from bias arising from the choice of calibration material and instrument, the carbon isotope ratios of steroids in urine extracts were measured using two different instruments in different laboratories, and three different reference materials (CU/USADA steroid standards from Brenna Laboratory, Cornell University; NIST RM8539 mineral oil; methane calibrated against NIST RM8560 natural gas). The measurements were performed at LGC and the Australian National Measurement Institute (NMI). It was found that there was no significant difference in measurement results when different instruments and reference materials were used to measure the carbon isotope ratio of the major testosterone metabolites androsterone and etiocholanolone, or the endogenous reference compounds pregnanediol, 11-ketoetiocholanolone and 11 beta-hydroxyandrosterone. Expanded measurement uncertainties at the 95% coverage probability ranged from 0.21 parts per thousand to 1.4 parts per thousand, depending on analyte, instrument and reference material. The measurement results of this comparison were used to estimate a measurement uncertainty of delta(13)C for the certification of the urine reference material being performed on a single instrument using a single reference material at NMI. Copyright (C) 2011 John Wiley & Sons, Ltd.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available