Discrepancies between isotope ratio infrared spectroscopy and isotope ratio mass spectrometry for the stable isotope analysis of plant and soil waters

Journal

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
Volume 24, Issue 14, Pages 1948-1954

Publisher

JOHN WILEY & SONS LTD
DOI: 10.1002/rcm.4597

Keywords

-

Funding

  1. Andrew W. Mellon Foundation
  2. NSF
  3. Center for Stable Isotope Biogeochemistry at UC Berkeley

Ask authors/readers for more resources

The use of isotope ratio infrared spectroscopy (IRIS) for the stable hydrogen and oxygen isotope analysis of water is increasing. While IRIS has many advantages over traditional isotope ratio mass spectrometry (IRMS), it may also be prone to errors that do not impact upon IRMS analyses. Of particular concern is the potential for contaminants in the water sample to interfere with the spectroscopy, thus leading to erroneous stable isotope data. Water extracted from plant and soil samples may often contain organic contaminants. The extent to which contaminants may interfere with IRIS and thus impact upon data quality is presently unknown. We tested the performance of IRIS relative to IRMS for water extracted from 11 plant species and one organic soil horizon. IRIS deviated considerably from IRMS for over half of the samples tested, with deviations as large as 46 parts per thousand (delta(2)H) and 15.4 parts per thousand (delta(18)O) being measured. This effect was reduced somewhat by using activated charcoal to remove organics from the water; however, deviations as large as 35 parts per thousand (delta(2)H) and 11.8 parts per thousand (delta(18)O) were still measured for these cleaned samples. Interestingly, the use of activated charcoal to clean water samples had less effect than previously thought for IRMS analyses. Our data show that extreme caution is required when using IRIS to analyse water samples that may contain organic contaminants. We suggest that the development of new cleaning techniques for removing organic contaminants together with instrument-based software to flag potentially problematic samples are necessary to ensure accurate plant and soil water analyses using IRIS. Copyright (C) 2010 John Wiley & Sons, Ltd.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available