The aim of this work was to develop and test a remote end-to-end audit system using lithium formate EPR dosimeters. Four clinics were included in a pilot study, absorbed doses determined in the PTV agreed with TPS calculated doses within +/- 5% for 3D-CRT and +/- 7% (kappa = 1) for IMRT/VMAT dose plans. (C) 2014 Elsevier Ireland Ltd. All rights reserved.